Brushes
Sample Holders ( chucks ) :
Put into box to "pre-chill"
Cryostat is kept at
Sectioning Device
Single Use Cryostat Blades
Optimal Cutting Temperature Compound ( O.C.T. )
Put on top of tissue sample , which is on top of chuck :
Turns white as it freezes :
Slides :
Roll Guard :
Check positioning of sample holder
should be set to around
each mark =
Blade Holder Lock - Forward / Back
You want to bring the whole apparatus close to where the tissue is on the sectioning device
Orange Lock = allows for pivoting of sectioning device
Normally , you don't need to mess with this
Open "plate" so you can insert the Roll Guard ( glass piece )
Make sure grooves are facing down
this step should be done early in the whole setup sequence , because the roll plate starts at room temperature , and needs time to cool down
Verify tissue sample is frozen ( O.C.T should be solid white )
Mount tissue sample into sectioning device :
Tighten Pivot Knob to keep sample from rotating
Get a blade out of the blade dispenser
Pull black slider piece with arrow all the way back
Then push black arrow piece slowly forward , should expose a single blade
Attach blade to cryostat
there is a tiny groove that it drops into
make sure it is centered ( left and right )
make sure it is not off at a weird angle ( should be flush with roll guard plate )
Use right-side orange locking handle to clamp the blade into place