Brushes
Sample Holders ( chucks ) :

Put into box to "pre-chill"

Cryostat is kept at

Sectioning Device

Single Use Cryostat Blades

Optimal Cutting Temperature Compound ( O.C.T. )

Put on top of tissue sample , which is on top of chuck :

Turns white as it freezes :

Slides :

Roll Guard :

Check positioning of sample holder
should be set to around
each mark =

Blade Holder Lock - Forward / Back
You want to bring the whole apparatus close to where the tissue is on the sectioning device

Orange Lock = allows for pivoting of sectioning device
Normally , you don't need to mess with this

Open "plate" so you can insert the Roll Guard ( glass piece )
Make sure grooves are facing down
this step should be done early in the whole setup sequence , because the roll plate starts at room temperature , and needs time to cool down


Verify tissue sample is frozen ( O.C.T should be solid white )

Mount tissue sample into sectioning device :

Tighten Pivot Knob to keep sample from rotating

Get a blade out of the blade dispenser
Pull black slider piece with arrow all the way back
Then push black arrow piece slowly forward , should expose a single blade

Attach blade to cryostat
there is a tiny groove that it drops into
make sure it is centered ( left and right )
make sure it is not off at a weird angle ( should be flush with roll guard plate )

Use right-side orange locking handle to clamp the blade into place