Upright microscope =
Week 1 = brightfield imaging on nervous system tissue
Week 2 = fluorescent imaging on nervous system tissue
Turn on the light source
Two modes for brightness
when illuminated it goes to a "preset"
if you click it again , you can then use the light intensity knob to manually adjust
just use the preset option
Make sure filters are correct
push button filters
all of them should be "out"
with the exception of light-balancing-filter ( LBD ) , should be pushed-in
Adjust "light path" knob
Pushed-In = light is only going to the eye piece
Half-Way-Pushed-In-Half-Way-Pulled-Out = divides light between eye piece and computer camera
you CAN use this for the brightfield pictures , but NOT the fluorescent ones next week
Pulled-Out = light is only going to the computer camera
Always start with a low magnification objective
Adjust Filter Wheel to position 1 for brightfield observation
Put the slide on the stage , spring
Translate stage as needed
Bottom = Left / Right Adjustment :
Top = Forward / Backward Adjustment
Bring Image into focus
Fine Focus = most lateral knob
Coarse Focus = most medial knob
Check that eye-pieces are set appropriately
Bottom = Width Adjustment
Left Eye-piece has Diopter Adjustment Ring = can adjust the focus of the left eye independent of what you see in the right eye
Start with slide on the specimen holder and with 10x objective lens
Look through eye pieces and confirm everything is in sharp focus
Adjust the condenser
using the condenser-height-adjustment-knob
Ok , so move the knob all the way to the "top"
aka turn all the way clockwise
Adjust field iris diaphragm
Adjust such that instead of a full beam of light , there is instead only a pin-hole amount
looking through the eyepieces as you do this should enable you to confirm its at a pin-hole amount
can confirm on computer camera screen , its only a small part of field of view
edges might not be perfectly sharp
should look like a stop sign when you are done
adjust the condenser hight knob until the edges look sharp
Now go back to the condenser , and make sure the spot is in the center of the field of view
This is done by adjusting two different centering screws underneath the condenser
on the left
and right side of the microscope
by turning the screws simultaneously , either loosening or tightening them
Now gradually open the field iris , until the light on the edges just barley goes beyond the field of view
easier to tell using the eye pieces
Example of "almost there"
So open it just a tad bit more ,
Adjust aperture iris slider
All the way to the right = very small spot of light that is illuminating the sample
All the way to the left = large are area of light illuminating the sample
0.9 - 0.2
Adjust based on the objective you are using and its numerical aperture
Example , we start with the 10x objective , with a 0.3 numerical aperature
so move the slider to match the numerical aperture of 0.3
If you switch objectives , you HAVE to adjust this slider every time to match
Green = 20x , with a 0.75 numerical aperature
So change slider to : 0.2 + around 5 lines = ~ 0.7
Set white balance
move sample to big open space where you have no tissue that the light is passing through
you want to tell the software that this has no color , so that it adjust the rest of the image to match this
Click top left button :
Then go over to the viewing window , and left click the mouse and drag to draw a box on region of interest ( ROI )
this re-calibrates the software to define this area as "white"
this gives good color representation for the rest of the image
When you are ready to save an image , click the "Snapshot" button
It asks you to confirm which objectives you are using
calibrations are already loaded into the computer
Click "OK"
It then opens another window with the "saved image"
one on the right should be the one you just "saved"
Now with the correct tab selected / highlighted , you can press the "Save" button or File -> Save As