When You First Get to the Microscope

  1. Turn on the light source

  2. Two modes for brightness

    • when illuminated it goes to a "preset"

      • already set to "mid-range" intensity
    • if you click it again , you can then use the light intensity knob to manually adjust

    • just use the preset option

      • gives uniform brightness for all samples
  3. Make sure filters are correct

    • push button filters

    • all of them should be "out"

      • with the exception of light-balancing-filter ( LBD ) , should be pushed-in

        • mimics "daylight" , even tone. instead of "yellowish light" that tungsten bulbs produce

  4. Adjust "light path" knob

    • Pushed-In = light is only going to the eye piece

    • Half-Way-Pushed-In-Half-Way-Pulled-Out = divides light between eye piece and computer camera

      • you CAN use this for the brightfield pictures , but NOT the fluorescent ones next week

        • fluorescent microscopy , you want to maximize the amount of light you are capturing , so pull it all the way out
    • Pulled-Out = light is only going to the computer camera

  5. Always start with a low magnification objective

    • Red Band = 4x
    • Yellow Band = 10x
    • Green Band = 20x
    • White Band = ? x
    • However , just start with the 10x , yellow band for our purposes
  6. Adjust Filter Wheel to position 1 for brightfield observation

  7. Put the slide on the stage , spring

  8. Translate stage as needed

    • Bottom = Left / Right Adjustment :

    • Top = Forward / Backward Adjustment

  9. Bring Image into focus

    • Fine Focus = most lateral knob

      • should only need fine focus for this assignment
    • Coarse Focus = most medial knob

  10. Check that eye-pieces are set appropriately

    • Bottom = Width Adjustment

    • Left Eye-piece has Diopter Adjustment Ring = can adjust the focus of the left eye independent of what you see in the right eye

      • useful because people's eyes vary , left eye vision is different than right
      • First close left eye , and focus on the sample using only your right eye
      • Then close right eye , and use ring to adjust viewing for left eye

How to Optimize the Illumination

  1. Start with slide on the specimen holder and with 10x objective lens

  2. Look through eye pieces and confirm everything is in sharp focus

  3. Adjust the condenser

    • using the condenser-height-adjustment-knob

      • up and down adjustment , focuses the lens of the condenser so that it matches the light path and the focal plane of the objectives

    • Ok , so move the knob all the way to the "top"

      • aka turn all the way clockwise

        • it might feel like its going to hit your slide and break it , but it won't
  4. Adjust field iris diaphragm

    • Counter-clockwise = "closes down"
    • Clockwise = "opens up"

    • Adjust such that instead of a full beam of light , there is instead only a pin-hole amount

    • looking through the eyepieces as you do this should enable you to confirm its at a pin-hole amount

    • can confirm on computer camera screen , its only a small part of field of view

    • edges might not be perfectly sharp

    • should look like a stop sign when you are done

    • adjust the condenser hight knob until the edges look sharp

  5. Now go back to the condenser , and make sure the spot is in the center of the field of view

    • This is done by adjusting two different centering screws underneath the condenser

      • on the left

      • and right side of the microscope

      • by turning the screws simultaneously , either loosening or tightening them

        • you can "walk" the spot of light around on the image you see on the computer screen
        • you want the spot of light in the middle of the field of view
  6. Now gradually open the field iris , until the light on the edges just barley goes beyond the field of view

    • easier to tell using the eye pieces

    • Example of "almost there"

    • So open it just a tad bit more ,

  7. Adjust aperture iris slider

    • All the way to the right = very small spot of light that is illuminating the sample

    • All the way to the left = large are area of light illuminating the sample

    • 0.9 - 0.2

      • lines = 0.1
    • Adjust based on the objective you are using and its numerical aperture

    • Example , we start with the 10x objective , with a 0.3 numerical aperature

    • so move the slider to match the numerical aperture of 0.3

      • 0.2 + 1 line = 0.3

    • If you switch objectives , you HAVE to adjust this slider every time to match

      • Green = 20x , with a 0.75 numerical aperature

      • So change slider to : 0.2 + around 5 lines = ~ 0.7

      • Also since you just changed the objective , change the field iris again to make sure the illuminated space is "just bigger" than the field of view you can see

 

CellSens Software

  1. Set white balance

    • move sample to big open space where you have no tissue that the light is passing through

    • you want to tell the software that this has no color , so that it adjust the rest of the image to match this

    • Click top left button :

    • Then go over to the viewing window , and left click the mouse and drag to draw a box on region of interest ( ROI )

    • this re-calibrates the software to define this area as "white"

    • this gives good color representation for the rest of the image

  2. When you are ready to save an image , click the "Snapshot" button

    • It asks you to confirm which objectives you are using

      • calibrations are already loaded into the computer

    • Click "OK"

    • It then opens another window with the "saved image"

      • one on the right should be the one you just "saved"

    • Now with the correct tab selected / highlighted , you can press the "Save" button or File -> Save As